Book a Demo!
CoCalc Logo Icon
StoreFeaturesDocsShareSupportNewsAboutPoliciesSign UpSign In
Download
29548 views
1
2
3
4
5
Methods for Measuring the Acute Toxicity of Effluents and
6
Receiving Waters to Freshwater and Marine Organisms
7
Fifth Edition
8
October 2002
9
10
11
12
U.S. Environmental Protection AgencyOffice of Water (4303T)
13
1200 Pennsylvania Avenue, NW Washington, DC 20460
14
EPA-821-R-02-012
15
The Engineering and Analysis Division, of the Office of Science
16
and Technology, has reviewed and approved this report for
17
publication. Neither the United States Government nor any of its
18
employees, contractors, or their employees make any warranty,
19
expressed or implied, or assumes any legal liability or
20
responsibility for any third party's use of or the results of such
21
use of any information, apparatus, product, or process discussed in
22
this report, or represents that its use by such party would not
23
infringe on privately owned rights.
24
25
INTRODUCTION
26
27
28
1.1
29
This manual describes acute toxicity tests for use in the
30
National Pollutant Discharge Elimination System (NPDES) Permits
31
Program to identify effluents and receiving waters containing toxic
32
materials in acutely toxic concentrations. With the exception of
33
the Holmesimysis costata Acute Test (Table 19), the methods
34
included in this manual are referenced in Table IA, 40 CFR Part 136
35
regulations and, therefore, constitute approved methods for acute
36
toxicity tests. They are also suitable for determining the toxicity
37
of specific compounds contained in discharges. The tests may be
38
conducted in a central laboratory or on-site, by the regulatory
39
agency or the permittee. The Holmesimysis costata Acute Test (Table
40
19) is specific to Pacific Coast waters and is not listed at 40 CFR
41
Part 136 for nationwide use. This method has been proposed but not
42
yet approved at 40 CFR Part 136.
43
44
45
1.2
46
The data are used for NPDES permits development and to
47
determine compliance with permit toxicity limits. Data can also be
48
used to predict potential acute and chronic toxicity in the
49
receiving water, based on the LC50 and appropriate dilution,
50
application, and persistence factors. The tests are performed as a
51
part of self-monitoring permit requirements, compliance
52
biomonitoring inspections, toxics sampling inspections, and special
53
investigations. Data from acute toxicity tests performed as part of
54
permit requirements are evaluated during compliance evaluation
55
inspections and performance audit inspections.
56
57
58
1.3
59
Modifications of these tests are also used in toxicity
60
reduction evaluations and toxicity identification evaluations to
61
identify the toxic components of an effluent, to aid in the
62
development and implementation of toxicity reduction plans, and to
63
compare and control the effectiveness of various treatment
64
technologies for a given type of industry, irrespective of the
65
receiving water (USEPA, 1988a; USEPA, 1988b; USEPA, 1989a; USEPA,
66
1989b; USEPA, 1991a).
67
68
69
1.4
70
This methods manual serves as a companion to the
71
short-term chronic toxicity test methods manuals for freshwater and
72
marine organisms (USEPA, 2002a; USEPA, 2002b), the NPDES compliance
73
inspection manual (USEPA, 1988c), and the manual for evaluation of
74
laboratories performing aquatic toxicity tests (USEPA, 1991b). In
75
2002, EPA revised previous editions of each of the three methods
76
manuals (USEPA, 1993a; USEPA, 1994a; USEPA, 1994b).
77
78
79
1.5
80
Guidance for the implementation of toxicity tests in the
81
NPDES program is provided in the Technical Support Document for
82
Water Quality-based Toxics Control (USEPA, 1991c).
83
84
85
1.6
86
The use of any test species or test conditions other than
87
those described in Tables 12-18 in this manual and referenced in
88
Table 1A, 40 CFR 136.3, shall be considered a major modification to
89
the method and subject to application and approval of alternate
90
test procedures under 40 CFR 136.4 and 40 CFR 136.5.
91
92
93
1.7
94
These methods are restricted to use by, or under the
95
supervision of, analysts experience in the use or conduct of, and
96
interpretation of data from, aquatic toxicity tests. Each analyst
97
must demonstrate the ability to generate acceptable test results
98
with the methods using the procedures described in this methods
99
manual.
100
101
102
1.8
103
This manual was prepared in the established
104
EMSL-Cincinnati format (USEPA, 1983a).
105
106
107
2.1
108
The selection of the test type will depend on the NPDES
109
permit requirements, the objectives of the test, the available
110
resources, the requirements of the test organisms, and effluent
111
characteristics such as fluctuations in effluent
112
toxicity.
113
114
115
2.2
116
Effluent acute toxicity is generally measured using a
117
multi-concentration, or definitive test, consisting of a control
118
and a minimum of five effluent concentrations. The tests are
119
designed to provide dose-response information, expressed as the
120
percent effluent concentration that is lethal to 50% of the test
121
organisms (LC50) within the prescribed period of time (24-96 h), or
122
the highest effluent concentration in which survival is not
123
statistically significantly different from the control.
124
125
126
2.3
127
Use of pass/fail tests consisting of a single effluent
128
concentration (e.g., the receiving water concentration or RWC) and
129
a control is not recommended. If the NPDES permit has a whole
130
effluent toxicity limit for acute toxicity at the RWC, it is
131
prudent to use that permit limit as the midpoint of a series of
132
five effluent concentrations. This will ensure that there is
133
sufficient information on the dose-response relationship. For
134
example, the effluent concentrations utilized in a test may be: (1)
135
100% effluent, (2) (RWC + 100)/2, (3) RWC, (4) RWC/2, and (5)
136
RWC/4. More specifically, if the RWC = 50%, appropriate effluent
137
concentrations may be 100%, 75%, 50%, 25%, and 12.5%.
138
139
140
2.4
141
Receiving (ambient) water toxicity tests commonly employ
142
two treatments, a control and the undiluted receiving water, but
143
may also consist of a series of receiving water
144
dilutions.
145
146
147
2.5
148
A negative result from an acute toxicity test does not
149
preclude the presence of chronic toxicity. Also, because of the
150
potential temporal variability in the toxicity of effluents, a
151
negative test result with a particular sample does not preclude the
152
possibility that samples collected at some other time might exhibit
153
acute (or chronic) toxicity.
154
155
156
2.6
157
The frequency with which acute toxicity tests are
158
conducted under a given NPDES permit is determined by the
159
regulatory agency on the basis of factors such as the variability
160
and degree of toxicity of the waste, production schedules, and
161
process changes.
162
163
164
2.7
165
Tests may be static (static non-renewal or static
166
renewal), or flow-through.
167
168
169
2.7.1 STATIC TESTS
170
171
172
2.7.1.1
173
Static non-renewal tests - The test organisms are exposed
174
to the same test solution for the duration of the test.
175
176
177
2.7.1.2
178
Static-renewal tests - The test organisms are exposed to
179
a fresh solution of the same concentration of sample every 24 h or
180
other prescribed interval, either by transferring the test
181
organisms from one test chamber to another, or by replacing all or
182
a portion of solution in the test chambers.
183
184
185
2.7.2
186
FLOW-THROUGH TESTS
187
188
189
190
191
2.7.2.1
192
Two types of flow-through tests are in common use: (1)
193
sample is pumped continuously from the sampling point directly to
194
the dilutor system; and (2) grab or composite samples are collected
195
periodically, placed in a tank adjacent to the test laboratory, and
196
pumped continuously from the tank to the dilutor system. The
197
flow-through method employing continuous sampling is the preferred
198
method for on-site tests. Because of the large volume (often 400
199
L/day) of effluent normally required for flow-through tests, it is
200
generally considered too costly and impractical to conduct these
201
tests off-site at a central laboratory.
202
203
204
2.8
205
Advantages and disadvantages of the types of tests are as
206
follows:
207
208
209
2.8.1 STATIC NON-RENEWAL TESTS
210
2.8.1.1 Advantages:
211
212
213
1.
214
Simple and inexpensive.
215
216
217
2.
218
Very cost effective in determining compliance with permit
219
conditions.
220
221
222
3.
223
Limited resources (space, manpower, equipment) required;
224
would permit staff to perform many more tests in the same amount of
225
time.
226
227
228
4.
229
Smaller volume of effluent required than for static
230
renewal or flow-through tests.
231
232
233
2.8.1.2 Disadvantages:
234
235
236
1.
237
Dissolved oxygen (DO) depletion may result from high
238
chemical oxygen demand (COD), biological oxygen demand (BOD), or
239
metabolic wastes.
240
241
242
2.
243
Possible loss of toxicants through volatilization and/or
244
adsorption to the exposure vessels.
245
246
247
3.
248
Generally less sensitive than static renewal or
249
flow-through tests, because the toxic substances may degrade or be
250
adsorbed, thereby reducing the apparent toxicity. Also, there is
251
less chance of detecting slugs of toxic wastes, or other temporal
252
variations in waste properties.
253
254
255
2.8.2 STATIC-RENEWAL, ACUTE TOXICITY TESTS
256
2.8.2.1 Advantages:
257
258
259
1.
260
Reduced possibility of dissolved oxygen (DO) depletion
261
from high chemical oxygen demand (COD) and/or biological oxygen
262
demand (BOD), or ill effects from metabolic wastes from organisms
263
in the test solutions.
264
265
266
2.
267
Reduced possibility of loss of toxicants through
268
volatilization and/or adsorption to the exposure
269
vessels.
270
271
272
3.
273
Test organisms that rapidly deplete energy reserves are
274
fed when the test solutions are renewed, and are maintained in a
275
healthier state.
276
277
278
2.8.2.2 Disadvantages:
279
280
281
1.
282
Require greater volume of effluent that non-renewal
283
tests.
284
285
286
2.
287
Generally less sensitive than flow-through tests, because
288
the toxic substances may degrade or be adsorbed, thereby reducing
289
the apparent toxicity. Also, there is less chance of detecting
290
slugs of toxic wastes, or other temporal variations in waste
291
properties.
292
293
294
2.8.3 FLOW-THROUGH TESTS
295
2.8.3.1 Advantages:
296
297
298
1.
299
Provide a more representative evaluation of the acute
300
toxicity of the source, especially if sample is pumped continuously
301
directly from the source and its toxicity varies with
302
time.
303
304
305
2.
306
DO concentrations are more easily maintained in the test
307
chambers.
308
309
310
3.
311
A higher loading factor (biomass) may be used.
312
313
314
4.
315
The possibility of loss of toxicant due to
316
volatilization, adsorption, degradation, and uptake is
317
reduced.
318
319
320
2.8.3.2 Disadvantages:
321
322
323
1.
324
Large volumes of sample and dilution water are
325
required.
326
327
328
2.
329
Test equipment is more complex and expensive, and
330
requires more maintenance and attention.
331
332
333
3.
334
More space is required to conduct tests.
335
336
337
4.
338
Because of the resources required, it would be very
339
difficult to perform multiple or overlapping sequential
340
tests.
341
342
343
HEALTH AND SAFETY
344
3.1 GENERAL PRECAUTIONS
345
346
347
3.1.1
348
Development and maintenance of an effective health and
349
safety program in the laboratory requires an ongoing commitment by
350
laboratory management, and includes (1) the appointment of a
351
laboratory health and safety officer with the responsibility and
352
authority to develop and maintain a safety program, (2) the
353
preparation of a formal, written, health and safety plan, which is
354
provided to each laboratory staff member, (3) an ongoing training
355
program on laboratory safety, and (4) regularly scheduled,
356
documented, safety inspections.
357
358
359
3.1.2
360
Collection and use of effluents in toxicity tests may
361
involve significant risks to personal safety and health. Personnel
362
collecting effluent samples and conducting toxicity tests should
363
take all safety precautions necessary for the prevention of bodily
364
injury and illness which might result from ingestion or invasion of
365
infectious agents, inhalation or absorption of corrosive or toxic
366
substances through skin contact, and asphyxiation due to lack of
367
oxygen or presence of noxious gases.
368
369
370
3.1.3
371
Prior to sample collection and laboratory work, personnel
372
must determine that all required safety equipment and materials
373
have been obtained and are in good condition.
374
375
376
3.1.4
377
Guidelines for the handling and disposal of hazardous
378
materials must be strictly followed.
379
380
381
3.2
382
SAFETY EQUIPMENT
383
384
385
3.2.1 PERSONAL SAFETY GEAR
386
387
388
3.2.1.1
389
Personnel must use safety equipment, as required, such as
390
rubber aprons, laboratory coats, respirators, gloves, safety
391
glasses, hard hats, and safety shoes.
392
393
394
3.2.2
395
LABORATORY SAFETY EQUIPMENT
396
397
398
399
400
3.2.2.1
401
Each laboratory (including mobile laboratories) must be
402
provided with safety equipment such as first aid kits, fire
403
extinguishers, fire blankets, emergency showers, and eye
404
fountains.
405
406
407
3.2.2.2
408
Mobile laboratories should be equipped with a telephone
409
to enable personnel to summon help in case of emergency.
410
411
412
3.3
413
GENERAL LABORATORY AND FIELD OPERATIONS
414
415
416
417
418
3.3.1
419
Guidance in Material Safety Data Sheets should be
420
followed for reagents and other chemicals purchased from supply
421
houses. Incompatible materials should not be stored
422
together.
423
424
425
3.3.2
426
Work with effluents must be performed in compliance with
427
accepted rules pertaining to the handling of hazardous materials
428
(see Safety Manuals, Subsection 3.5). Personnel collecting samples
429
and performing toxicity tests should not work alone.
430
431
432
3.3.3
433
Because the chemical composition of effluents is usually
434
only poorly known, they must be considered as potential health
435
hazards, and exposure to them should be minimized. Fume and canopy
436
hoods over the test areas must be used whenever
437
necessary.
438
439
440
3.3.4
441
It is advisable to cleanse exposed parts of the body
442
immediately after collecting effluent samples.
443
444
445
3.3.5
446
All containers must be adequately labeled to indicate
447
their contents.
448
449
450
3.3.6
451
Strong acids and volatile organic solvents employed in
452
glassware cleaning must be used in a fume hood or under an exhaust
453
canopy over the work area.
454
455
456
3.3.7
457
Good housekeeping contributes to safety and reliable
458
results.
459
460
461
3.3.8
462
Electrical equipment or extension cords not bearing the
463
approval of Underwriter Laboratories must not be used. Ground-fault
464
interrupters must be installed in all "wet" laboratories where
465
electrical equipment is used.
466
467
468
3.3.9
469
Mobile laboratories must be properly grounded to protect
470
against electrical shock.
471
472
473
3.4
474
DISEASE PREVENTION
475
476
477
478
479
3.4.1
480
Personnel handling samples which are known or suspected
481
to contain human wastes should be immunized against hepatitis B,
482
tetanus, typhoid fever, and polio.
483
484
485
3.5
486
SAFETY MANUALS
487
488
489
490
491
3.5.1
492
For further guidance on safe practices when collecting
493
effluent samples and conducting toxicity tests, check with the
494
permittee and consult general industrial safety manuals, including
495
USEPA (1986) and Walters and Jameson (1984).
496
497
498
3.6
499
WASTE DISPOSAL
500
501
502
3.6.1 Wastes generated during toxicity testing must be properly
503
handled and disposed of in an appropriate manner. Each testing
504
facility will have its own waste disposal requirements based on
505
local, state, and Federal rules and regulations. It is extremely
506
important that these rules and regulations be known, understood,
507
and complied with by all persons responsible for, or otherwise
508
involved in, performing testing activities. Local fire officials
509
should be notified of any potentially hazardous conditions.
510
QUALITY ASSURANCE
511
4.1 INTRODUCTION
512
4.1.1 Development and maintenance of a toxicity test laboratory
513
quality assurance (QA) program requires an ongoing commitment by
514
laboratory management, and includes the following: (1) appointment
515
of a laboratory quality assurance officer with the responsibility
516
and authority to develop and maintain a QA program;
517
(2) preparation of a quality assurance plan with data quality
518
objectives; (3) preparation of written descriptions of laboratory
519
standard operating procedures (SOP's) for test organism culturing,
520
toxicity testing, instrument calibration, sample chain-of-custody,
521
laboratory sample tracking system, etc.; and (4) provision of
522
adequate, qualified technical staff and suitable space and
523
equipment to assure reliable data.
524
4.1.2 QA practices within an aquatic toxicology laboratory must
525
address all activities that affect the quality of the final
526
effluent toxicity data, such as: (1) effluent sampling and
527
handling; (2) the source and condition of the test organisms; (3)
528
condition and operation of equipment; (4) test conditions; (5)
529
instrument calibration; (6) replication;
530
(7) use of reference toxicants; (8) record keeping; and (9) data
531
evaluation.
532
533
534
4.1.3
535
Quality control practices, on the other hand, consist of
536
the more focused, routine, day-to-day activities carried out within
537
the scope of the overall QA program. For more detailed discussion
538
of quality assurance, and general guidance on good laboratory
539
practices related to toxicity testing, see: FDA, 1978; USEPA, 1975;
540
USEPA, 1979a; USEPA, 1980a; USEPA, 1980b; USEPA, 1991b; DeWoskin,
541
1984; and Taylor, 1987.
542
543
544
4.1.4
545
Guidance for the evaluation of laboratories performing
546
toxicity tests and laboratory evaluation criteria may be found in
547
USEPA 1991b.
548
549
550
4.2
551
FACILITIES, EQUIPMENT, AND TEST CHAMBERS
552
553
554
555
556
4.2.1
557
Separate test organism culturing and toxicity testing
558
areas should be provided to avoid possible loss of cultures due to
559
cross-contamination. Ventilation systems should be designed and
560
operated to prevent recirculation or leakage of air from chemical
561
analysis laboratories or sample storage and preparation areas into
562
organism culturing or toxicity testing areas, and from toxicity
563
test laboratories and sample preparation areas into culture
564
rooms.
565
566
567
4.2.2
568
Laboratory and toxicity test
569
temperature control equipment must be adequate to maintain
570
recommended test water temperatures. Recommended materials must be
571
used in the fabrication of the test equipment which comes in
572
contact with the effluent (see Section 5, Facilities and
573
Equipment).
574
575
576
4.3
577
TEST ORGANISMS
578
579
580
581
582
4.3.1
583
The test organisms used in the procedures described in
584
this manual are listed in Section 6, Test Organisms. The organisms
585
should appear healthy, behave normally, feed well, and have low
586
mortality in cultures, during holding, and in test controls. Test
587
organisms should be positively identified to species.
588
589
590
4.4
591
LABORATORY WATER USED FOR CULTURING AND TEST DILUTION
592
WATER
593
594
595
4.4.1 The quality of water used for test organism culturing and
596
for dilution water used in toxicity tests is extremely important.
597
Water for these two uses should come from the same source. The
598
dilution water used in effluent toxicity tests will depend in part
599
on the objectives of the study and logistical constraints, as
600
discussed in detail in Section 7, Dilution Water. The dilution
601
water used for internal quality assurance tests with organisms,
602
food, and reference toxicants should be the water routinely used
603
with success in the laboratory. Types of water are discussed in
604
Section 5, Facilities and Equipment. Water used for culturing and
605
test dilution should be analyzed for toxic metals and organics at
606
least annually or whenever difficulty is encountered in meeting
607
minimum acceptability criteria for control survival and
608
reproduction or growth. The concentration of the metals, Al, As,
609
Cr, Co, Cu, Fe, Pb, Ni, and Zn, expressed as total metal, should
610
not exceed 1 µg/L each, and Cd, Hg, and Ag, expressed as total
611
metal, should not exceed 100 ng/L each. Total organochlorine
612
pesticides plus PCBs should be less than 50 ng/L (APHA, 1992).
613
Pesticide concentrations should not exceed USEPA's Ambient Water
614
Quality chronic criteria values where available.
615
4.5 EFFLUENT SAMPLING AND SAMPLE HANDLING
616
617
618
4.5.1
619
Sample holding times and temperatures must conform to
620
conditions described in Section 8, Effluent and Receiving Water
621
Sampling and Sample Handling.
622
623
624
4.6
625
TEST CONDITIONS
626
627
628
629
630
4.6.1
631
The temperature of test solutions
632
must be measured by placing the thermometer or probe directly into
633
the test solutions, or by placing the thermometer in equivalent
634
volumes of water in surrogate vessels positioned at appropriate
635
locations among the test vessels. Temperature should be recorded
636
continuously in at least one vessel during the duration of each
637
test. Test solution temperatures should be maintained within the
638
limits specified for each test. DO concentration and pH in test
639
chambers should be checked daily throughout the test period, as
640
prescribed in Section 9, Acute Toxicity Test Procedures.
641
642
643
4.7
644
QUALITY OF TEST ORGANISMS
645
646
647
648
649
4.7.1
650
The health of test organisms is primarily assessed by the
651
performance (survival, growth, and/or reproduction) of organisms in
652
control treatments of individual tests. The health and sensitivity
653
of test organisms is also assessed by reference toxicant testing.
654
In addition to documenting the sensitivity and health of test
655
organisms, reference toxicant testing is used to initially
656
demonstrate acceptable laboratory performance (Subsection 4.14) and
657
to document ongoing laboratory performance (Subsection
658
4.15).
659
660
661
4.7.2
662
Regardless of the source of test organisms (in-house
663
cultures or purchased from external suppliers), the testing
664
laboratory must perform at least one acceptable reference toxicant
665
test per month for each toxicity test method conducted in that
666
month (Subsection 4.15). If a test method is conducted only
667
monthly, or less frequently, a reference toxicant test must be
668
performed concurrently with each effluent toxicity test.
669
670
671
4.7.3
672
When acute or short-term chronic toxicity tests are
673
performed with effluents or receiving waters using test organisms
674
obtained from outside the test laboratory, concurrent toxicity
675
tests of the same type must be preformed with a reference toxicant,
676
unless the test organism supplier provides control chart data from
677
at least the last five monthly acute toxicity tests using the same
678
reference toxicant and test conditions.
679
680
681
4.7.4
682
The supplier should also certify the species
683
identification of the test organisms, and provide the taxonomic
684
reference (citation and page) or name(s) of the taxonomic expert(s)
685
consulted.
686
687
688
4.7.5
689
If a routine reference toxicant test fails to meet test
690
acceptability criteria, then the reference toxicant test must be
691
immediately repeated.
692
693
694
4.8
695
FOOD QUALITY
696
697
698
699
700
4.8.1
701
The nutritional quality of the food used in culturing and
702
testing fish and invertebrates is an important factor in the
703
quality of the toxicity test data. This is especially true for the
704
unsaturated fatty acid content of brine shrimp nauplii, Artemia.
705
Suitable trout chow, Artemia, and other foods must be obtained as
706
described in this manual.
707
708
709
4.8.2
710
Problems with the nutritional suitability of the food
711
will be reflected in the survival, growth, and reproduction of the
712
test organisms in cultures and toxicity tests. If a batch of food
713
is suspected to be defective, the performance of organisms fed with
714
the new food can be compared with the performance of organisms fed
715
with a food of known quality in side-by-side tests. If the food is
716
used for culturing, its suitability should be determined using a
717
short-term chronic test which will determine the effect of food
718
quality on growth or reproduction of each of the relevant test
719
species in culture, using four replicates with each food source.
720
Where applicable, foods used only in acute toxicity tests can be
721
compared with a food of known quality in side-by-side,
722
multi-concentration acute tests, using the reference toxicant
723
regularly employed in the laboratory QA program.
724
725
726
4.8.3
727
New batches of food used in culturing and testing should
728
be analyzed for toxic organics and metals or whenever difficulty is
729
encountered in meeting minimum test acceptability criteria for
730
control survival and reproduction or growth. If the concentration
731
of total organochlorine pesticides exceeds 0.l5 µg/g wet weight, or
732
the concentration of the total organochlorine pesticides plus PCBs
733
exceeds 0.30 µg/g wet weight, or toxic metals (Al, As, Cr, Co, Cu,
734
Pb, Ni, Zn, expressed as total metal) exceed 20 µg/g wet weight,
735
the food should not be used (for analytical methods see AOAC, 1990
736
and USDA, 1989). For foods (e.g., such as YCT) which are used to
737
culture and test organisms, the quality of food should meet the
738
requirements for the laboratory water used for culturing and test
739
dilution water as described in Section 4.4 above.
740
741
742
4.9
743
ACCEPTABILITY OF ACUTE TOXICITY TEST RESULTS
744
745
746
747
748
4.9.1
749
For the test results to be acceptable, control survival
750
must equal or exceed 90%.
751
752
753
4.9.2
754
An individual test may be conditionally acceptable if
755
temperature, DO, and other specified conditions fall outside
756
specifications, depending on the degree of the departure and the
757
objectives of the tests (see test condition summaries). The
758
acceptability of the test will depend on the experience and
759
professional judgment of the laboratory analyst and the reviewing
760
staff of the regulatory authority. Any deviation from test
761
specifications must be noted when reporting data from a
762
test.
763
764
765
4.10 ANALYTICAL METHODS
766
767
768
4.10.1
769
Routine chemical and physical analyses for culture and
770
dilution water, food, and test solutions, must include established
771
quality assurance practices outlined in Agency methods manuals
772
(USEPA, 1979a; USEPA, 1993b).
773
774
775
4.10.2
776
Reagent containers should be dated when received from the
777
supplier, and the shelf life should not be exceeded. Also, working
778
solutions should be dated when prepared, and the recommended shelf
779
life should be observed.
780
781
782
4.11
783
CALIBRATION AND STANDARDIZATION
784
785
786
787
788
4.11.1
789
Instruments used for routine measurements of chemical and
790
physical parameters such as pH, DO, temperature, conductivity,
791
salinity, alkalinity, and hardness must be calibrated and
792
standardized prior to use each day according to the instrument
793
manufacturer's procedures as indicated in the general section on
794
quality assurance (see EPA Methods 150.1, 360.1, 170.1, and 120.1;
795
USEPA, 1979b). Calibration data are recorded in a permanent
796
log.
797
798
799
4.11.2
800
Wet chemical methods used to measure hardness,
801
alkalinity, and total residual chlorine must be standardized prior
802
to use each day according to the procedures for those specific EPA
803
methods (see EPA Methods 130.2 and 310.1; USEPA 1979b).
804
805
806
4.12
807
REPLICATION AND TEST SENSITIVITY
808
809
810
811
812
4.12.1
813
The sensitivity of toxicity tests
814
will depend in part on the number of replicates per concentration,
815
the significance level selected, and the type of statistical
816
analysis. If the variability remains constant, the sensitivity of
817
the test will increase as the number of replicates is increased.
818
The minimum recommended number of replicates varies with the
819
objectives of the test and the statistical method used for analysis
820
of the data.
821
822
823
4.13
824
VARIABILITY IN TOXICITY TEST RESULTS
825
826
827
828
829
4.13.1
830
Factors which can affect test success and precision
831
include: the experience and skill of the laboratory analyst; test
832
organism age, condition, and sensitivity; dilution water quality;
833
temperature control; and the quality and quantity of food provided.
834
The results will depend upon the species used and the strain or
835
source of the test organisms, and test conditions such as
836
temperature, DO, food, and water quality. The repeatability or
837
precision of toxicity tests is also a function of the number of
838
test organisms used at each toxicant concentration. Jensen (1972)
839
discussed the relationship between sample size (numbers of fish)
840
and the standard error of the test, and considered 20 fish per
841
concentration as optimum for Probit Analysis.
842
843
844
4.13.2
845
Test precision can be estimated by using the same strain
846
of organisms under the same test conditions, and employing a known
847
toxicant, such as a reference toxicant. The single-laboratory
848
(intra-laboratory) and multi-laboratory (inter-laboratory)
849
precision of acute toxicity tests with several common test species
850
and reference toxicants are listed in Tables 1-4. Intra- and
851
inter-laboratory precision are described by the mean, standard
852
deviation, and relative standard deviation (percent coefficient of
853
variation, or CV) of the calculated endpoints from the replicated
854
tests.
855
856
857
4.13.3
858
Intra-laboratory precision data from 268 acute toxicity
859
tests with four species and five reference toxicants are listed in
860
Tables 1 and 2. The precision, expressed as CV%, ranged from 3% to
861
86%. More recent CV values reported by Jop et al. (1986), Dorn and
862
Rogers (1989), Hall et al. (1989), and Cowgill et al. (1990), fell
863
in a somewhat lower range (8% to 41%).
864
865
866
4.13.4
867
Inter-laboratory precision of acute toxicity tests from
868
253 reference toxicant tests with seven species, listed in Tables
869
2, 3, 4, and 5 (expressed as CV% for LC50s), ranged from 11% to
870
167%. Table 6 shows interlaboratory precision data from a study of
871
acute toxicity test methods using reference toxicant, effluent, and
872
receiving water sample types (USEPA, 2001a; USEPA, 2001b). Averaged
873
across sample types, total interlaboratory precision (expressed as
874
CV% for LC50s) ranged from 13% to 38.5% for the acute
875
methods.
876
877
878
4.13.5
879
No clear pattern of differences were noted in the intra-
880
or inter-laboratory test precision with the species listed,
881
although the test results with some toxicants, such as cadmium,
882
appear to more variable than those with other reference
883
toxicants.
884
885
886
4.13.6
887
Additional information on toxicity test precision is
888
provided in the Technical Support Document for Water Quality-Based
889
Toxics Control (see pp. 2-4, and 11-15; USEPA, 1991c).
890
891
892
893
1
894
Precision expressed as percent coefficient of variation, where
895
CV% = (standard deviation X 100)/mean.
896
2
897
SDS = Sodium dodecyl (lauryl) sulfate; NAPCP = Sodium
898
pentachlorophenate; CD = Cadmium; N = Number of tests; toxicant
899
concentration in mg/L.
900
3
901
Pimephales promelas tests were performed in soft, synthetic
902
freshwater; total hardness, 40-48 mg/L as CaCO3, by J. Dryer,
903
Aquatic Biology Section, EMSL-Cincinnati.
904
4
905
Daphnia data from Lewis and Horning, 1991. Tests with D. magna
906
used hard reconstituted water (total
907
hardness, 180-200 mg/L as CaCO3); tests with D. pulex used
908
moderately-hard reconstituted water (total
909
hardness, 80-100 mg/L as CaCO3).
910
5
911
Mysid tests were performed in 25 ppt salinity, natural seawater.
912
Data were provided by Steve Ward,
913
Environmental Services Division, U.S. Environmental Protection
914
Agency, Edison, New Jersey. Personal
915
communication, November 14, 1990.
916
917
918
1
919
From Table 2, p. 191, Grothe and Kimerle, 1985. Tests performed
920
at 20EC ±2EC; dilution water hardness, 100mg/L as CaCO3; dilution
921
water alkalinity, 76 mg/L as CaCO3; effluent hardness, approx. 1000
922
mg/L as CaCO3; effluent alkalinity, 310 mg/L as CaCO3; effluent
923
dilutions - 56%, 32%, 18%, 10%, 5.6%, 3.1%, 1.7%.
924
2
925
LC50 expressed in percent effluent.
926
3
927
Intra-laboratory precision expressed as the weighted mean
928
CV(%).
929
REFERENCE TOXICANT
930
931
1
932
Data for Pimephales promelas (fathead minnow), Oncorhyncus
933
mykiss (rainbow trout), and Daphnia magna were taken from USEPA,
934
1983b.
935
Data for, Mysidopsis bahia, and Cyprinodon variegatus
936
(sheepshead minnow) were taken from USEPA, 1981. Six laboratories
937
participated in each study. Test salinity was 28‰.
938
LC50s expressed in µg/L.
939
In the studies with the freshwater organisms, the water hardness
940
for five of the six laboratories ranged between 36 and 75 mg/L.
941
However, the water hardness for the sixth laboratory was 255 mg/L,
942
resulting in LC50 values for silver more than an order of magnitude
943
larger than for the other five. These values were rejected in
944
calculating the CV%. The mean weights of test fish were from
945
0.05-0.26 g for fathead minnows, and 0.22-1.32 g for rainbow trout.
946
Daphnia were #24-h old.
947
In studies with the marine organisms, only one LC50 (presumably
948
the combined LC50 from duplicate tests) was reported for each
949
toxicity test. LC50s for flow-through tests with Mysidopsis bahia
950
and Cyprinodon variegatus were calculated two different ways -- (1)
951
on the basis of the nominal toxicant concentrations (Nom), and (2)
952
on the basis of measured (Meas) toxicant concentrations. Test
953
organism age was #2 days for Mysidopsis bahia, and 28 days for
954
Cyprinodon variegatus. The salinity of test solutions was 28‰.
955
N, the total number of LC50 values used in calculating the CV(%)
956
varied with organism and toxicant because some data were rejected
957
due to water hardness, lack of concentration measurements, and/or
958
because some of the LC50s were not calculable.
959
2
960
CV% = Percent coefficient of variation = (standard deviation x
961
100)/mean.
962
TEST PRECISION (CV%)2
963
GRAPH3STAT4NO. LABS METHOD METHOD TOTAL5
964
SUBMITTING TEST TYPE VALID DATA N LC50 CV% N LC50 CV% N LC50
965
CV%
966
Pimephales promelas (96 h, 22EC)6 17 Pimephales promelas (24 h,
967
25EC)7 6 Ceriodaphnia dubia (48 h, 25EC)7 11 Mysidopsis bahia (96
968
h, 22EC)8 14
969
6 944 28.8 13 832 27.8 17 864 29.6 6 83211.5 6 83211.5 - - - 11
970
256 53.111 264 48.5 - - - 7 292 32.9 11 250 36.0 14 268 37.3
971
1
972
Interlaboratory study of toxicity test precision conducted in
973
1990 by the Environmental Monitoring Systems Laboratory -
974
Cincinnati, U.S. Environmental Protection Agency, Cincinnati, Ohio
975
45268, in cooperation with the states of New Jersey and North
976
Carolina, and the Office of Water Enforcement and Permits, U.S.
977
Environmental Protection Agency, Washington, DC.
978
2
979
Percent coefficient of variation = (standard deviation X
980
100)/mean. Calculated for LC50 from acute tests. LC50s expressed as
981
mg/L KCl added to the dilution water.
982
3
983
LC50 estimated by the Graphical Method.
984
4
985
LC50 estimated by Probit, Litchfield-Wilcoxon, or Trimmed
986
Spearman-Karber method.
987
5
988
LC50 usually reported for only one method of analysis for each
989
test. Where more than one LC50 was reported for a test, the lowest
990
value was used to calculate the statistics for "Total."
991
6
992
Data from the New Jersey Department of Environmental Protection:
993
static daily-renewal tests, using moderately- hard synthetic
994
freshwater.
995
7
996
Data from North Carolina certified laboratories: static
997
non-renewal tests, using moderately-hard reconstituted
998
freshwater.
999
8
1000
Data from the New Jersey Department of Environmental Protection:
1001
static daily-renewal tests, using 25 ppt salinity, FORTY FATHOMS®
1002
synthetic seawater.
1003
No. Labs Test Type Submitting Data LC50 CV%2
1004
Pimephales promelas (48 h, 25EC)3 Ceriodaphnia dubia (48 h,
1005
25EC)3 Mysidopsis bahia (48 h, 25EC)5 Menidia beryllina (48 h,
1006
25EC)5
1007
203 8964 28.6 171 4324 39.8 61 5324 30.1 39 1646 42.2
1008
1
1009
From a national study of interlaboratory precision of toxicity
1010
test data performed in 1991 by the Environmental Monitoring Systems
1011
Laboratory - Cincinnati, U.S. Environmental Protection Agency,
1012
Cincinnati, OH 45268. Participants included Federal, state, and
1013
private laboratories engaged in NPDES permit compliance monitoring.
1014
LC50s were estimated by the graphical or Spearman-Karber
1015
method.
1016
2
1017
Percent coefficient of variation = (standard deviation X
1018
100)/mean.
1019
3
1020
Static non-renewal tests, using moderately-hard synthetic
1021
freshwater (total hardness = 80-100 mg/L as CaCO3).
1022
4
1023
Expressed as mg KCl added per liter of dilution water.
1024
5
1025
Static non-renewal tests, using 30 ppt modified GP2 artificial
1026
seawater.
1027
6
1028
Expressed as Fg Cu++ added per liter of dilution water.
1029
TABLE 6. NATIONAL INTERLABORATORY STUDY OF ACUTE TOXICITY TEST
1030
PRECISION, 2000: PRECISION OF LC50 POINT ESTIMATES FOR REFERENCE
1031
TOXICANT, EFFLUENT, AND RECEIVING WATER SAMPLE TYPES1.
1032
CV (%)2
1033
Method Sample Type
1034
Within-lab3 Between-lab4 Total5
1035
Pimephales promelas KCl Municipal effluent Receiving water
1036
7.62 19.7 21.1 10.3 19.2 21.8 --17.2 Average 8.96 19.4 20.0
1037
Ceriodaphnia dubia KCl Municipal effluent Receiving water
1038
14.6 15.2 21.1 9.68 32.8 34.2 --31.8 Average 12.1 24.0 29.0
1039
Cyprinodon variegatus KCl Municipal effluent Receiving water
1040
--26.0 --19.4 --32.5 Average --26.0
1041
6
1042
Menidia beryllina CuSO4 Industrial effluent Receiving water
1043
---9.91 49.7 50.7
1044
--26.3 Average 9.91 49.7 38.5
1045
Holmesimysis costata7 Zn (48 h test) Zn (96 h test) Zn
1046
(interlaboratory trial 1) Zn (interlaboratory trial 2)
1047
Average
1048
19
1049
23 ---24 --1
1050
21 13
1051
1
1052
From EPA's WET Interlaboratory Variability Study (USEPA, 2001a;
1053
USEPA, 2001b).
1054
2
1055
CVs were calculated based on the within-laboratory component of
1056
variability, the between-laboratory component of variability, and
1057
total interlaboratory variability (including both within-laboratory
1058
and between-laboratory components). For the receiving water sample
1059
type, within-laboratory and between-laboratory components of
1060
variability could not be calculated since the study design did not
1061
provide within-laboratory replication for this sample type. The
1062
study design also did not provide within-laboratory replication for
1063
the Cyprinodon variegatus Acute Method.
1064
3
1065
The within-laboratory (intralaboratory) component of variability
1066
for duplicate samples tested at the same time in the same
1067
laboratory.
1068
4
1069
The between-laboratory component of variability for duplicate
1070
samples tested at different laboratories.
1071
5
1072
The total interlaboratory variability, including
1073
within-laboratory and between-laboratory components of variability.
1074
The total interlaboratory variability is synonymous with
1075
interlaboratory variability reported from other studies where
1076
individual variability components are not separated.
1077
6
1078
Precision estimates were not calculated for the reference
1079
toxicant sample type since the majority of results for this sample
1080
type were outside of the test concentration range (ie.,
1081
>100).
1082
7
1083
Holmesimysis costata Acute Test data were from Martin et al.
1084
(1989). Zn was tested in two intralaboratory trials and in two
1085
interlaboratory trials. Data from this study was only reported to
1086
two significant figures.
1087
4.14 DEMONSTRATING ACCEPTABLE LABORATORY
1088
PERFORMANCE
1089
1090
1091
4.14.1
1092
It is a laboratory's responsibility to demonstrate its
1093
ability to obtain consistent, precise results with reference
1094
toxicants before it performs toxicity tests with effluents for
1095
permit compliance purposes. To meet this requirement, the
1096
intra-laboratory precision, expressed as percent coefficient of
1097
variation (CV%), of each type of test to be used in a laboratory
1098
should be determined by performing five or more tests with
1099
different batches of test organisms, using the same reference
1100
toxicant, at the same concentrations, with the same test conditions
1101
(i.e., the same test duration, type of dilution water, age of test
1102
organisms, feeding, etc.), and same data analysis methods. A
1103
reference toxicant concentration series (0.5 or higher) should be
1104
selected that will consistently provide partial mortalities at two
1105
or more concentrations.
1106
1107
1108
4.15
1109
DOCUMENTING ONGOING LABORATORY PERFORMANCE
1110
1111
1112
1113
1114
4.15.1
1115
Satisfactory laboratory performance is demonstrated by
1116
performing at least one acceptable test per month with a reference
1117
toxicant for each toxicity test method conducted in the laboratory
1118
during that month. For a given test method, successive tests must
1119
be performed with the same reference toxicant, at the same
1120
concentrations, in the same dilution water, using the same data
1121
analysis methods. Precision may vary with the test species,
1122
reference toxicant, and type of test. Each laboratory's reference
1123
toxicity data will reflect conditions unique to that facility,
1124
including dilution water, culturing, and other variables; however,
1125
each laboratory's reference toxicity results should reflect good
1126
repeatability.
1127
1128
1129
4.15.2
1130
A control chart should be prepared for each combination
1131
of reference toxicant, test species, test condition, and endpoint.
1132
Toxicity endpoints from five or six tests are adequate for
1133
establishing the control charts. In this technique, a running plot
1134
is maintained for the toxicity values (Xi) from successive tests
1135
with a given reference toxicant (Figure 1), and endpoints (LC50s)
1136
are examined to determine if they are within prescribed limits. The
1137
types of control charts illustrated (see USEPA, 1979a) are used to
1138
evaluate the cumulative trend of results from a series of samples,
1139
thus reference toxicant test results should not be used as a de
1140
facto criterion for rejection of individual effluent or receiving
1141
water tests. The mean ( ¯ ) and upper and lower control limits
1142
(±2S) are re-
1143
1144
1145
X calculated with each successive test result. After two years
1146
of data collection, or a minimum of 20 data points, the control
1147
chart should be maintained using only the 20 most recent data
1148
points.
1149
1150
1151
4.15.3
1152
Laboratories should compare the calculated CV (i.e.,
1153
standard deviation / mean) of the LC50 for the 20 most recent data
1154
points to the distribution of laboratory CVs reported nationally
1155
for reference toxicant testing (Table 3-3 in USEPA, 2000b). If the
1156
calculated CV exceeds the 75th percentile of CVs reported
1157
nationally, the laboratory should use the 75th and 90th percentiles
1158
to calculate warning and control limits, respectively, and the
1159
laboratory should investigate options for reducing
1160
variability.
1161
1162
1163
4.15.4
1164
The outliers, which are values falling outside the upper
1165
and lower control limits, and trends of increasing or decreasing
1166
sensitivity, are readily identified. At the P0.05 probability
1167
level, one in 20 tests would be expected to fall outside of the
1168
control limits by chance alone. If more than one out of 20
1169
reference toxicant tests fall outside the control limits, the
1170
laboratory should investigate sources of variability, take
1171
corrective actions to reduce identified sources of variability, and
1172
perform an additional reference toxicant test during the same
1173
month. In those instances when the laboratory can document the
1174
cause for the outlier (e.g., operator error, culture health or test
1175
system failure), the outlier should be excluded from the future
1176
calculations of the control limits. If two or more consecutive
1177
tests do not fall within the control limits, the results must be
1178
explained and the reference toxicant test must be immediately
1179
repeated. Actions taken to correct the problem must be
1180
reported.
1181
1182
1183
4.15.5
1184
If the toxicity value from a given test with the
1185
reference toxicant falls well outside the expected range for the
1186
test organisms when using the standard dilution water, the
1187
laboratory should investigate sources of variability, take
1188
corrective actions to reduce identified sources of variability, and
1189
perform an additional reference toxicant test during the same
1190
month. Performance should improve with experience, and the control
1191
limits for point estimates should gradually narrow. However,
1192
control limits of ±2S, by definition, will be exceeded 5% of the
1193
time, regardless of how well a laboratory performs. Highly
1194
proficient laboratories which develop a very narrow control limit
1195
may be unfairly penalized if a test which falls just outside the
1196
control limits is rejected de facto. For this reason, the
1197
width
1198
1199
1200
of the control limits should be considered in determining
1201
whether or not a reference toxicant test result falls "well"
1202
outside the expected range. The width of the control limits may be
1203
evaluated by comparing the calculated CV (i.e., standard deviation
1204
/ mean) of the LC50 for the 20 most recent data points to the
1205
distribution of laboratory CVs reported nationally for reference
1206
toxicant testing (Table 3-3 in USEPA, 2000b). In determining
1207
whether or not a reference toxicant test result falls "well"
1208
outside the expected range, the result also may be compared with
1209
upper and lower bounds for ±3S, as any result outside these control
1210
limits would be expected to occur by chance only 1 out of 100 tests
1211
(Environment Canada, 1990). When a result from a reference toxicant
1212
test is outside the 99% confidence intervals, the laboratory must
1213
conduct an immediate investigation to assess the possible causes
1214
for the outlier.
1215
1216
1217
4.15.6
1218
Reference toxicant test results should not be used as a
1219
de facto criterion for rejection of individual effluent or
1220
receiving water tests. Reference toxicant testing is used for
1221
evaluating the health and sensitivity of organisms over time and
1222
for documenting initial and ongoing laboratory performance. While
1223
reference toxicant test results should not be used as a de facto
1224
criterion for test rejection, effluent and receiving water test
1225
results should be reviewed and interpreted in the light of
1226
reference toxicant test results. The reviewer should consider the
1227
degree to which the reference toxicant test result fell outside of
1228
control chart limits, the width of the limits, the direction of the
1229
deviation (toward increased test organism sensitivity or toward
1230
decreased test organism sensitivity), the test conditions of both
1231
the effluent test and the reference toxicant test, and the
1232
objective of the test.
1233
1234
1235
4.16
1236
REFERENCE TOXICANTS
1237
1238
1239
1240
1241
1242
4.16.1
1243
Reference toxicants such as sodium
1244
chloride (NaCl), potassium chloride (KCl), cadmium chloride
1245
(CdCl2), copper sulfate (CuSO4), sodium dodecyl sulfate (SDS), and
1246
potassium dichromate (K2Cr2O7), are suitable for use in the NPDES
1247
and other Agency programs requiring aquatic toxicity tests.
1248
EMSL-Cincinnati hopes to release EPAcertified solutions of cadmium
1249
and copper, with accompanying toxicity data for the recommended
1250
test species, for use as reference toxicants through cooperative
1251
research and development agreements with commercial suppliers, and
1252
will continue to develop additional reference toxicants for future
1253
release. Standard reference materials can be obtained from
1254
commercial supply houses, or can be prepared inhouse using reagent
1255
grade chemicals. The regulatory agency should be consulted before
1256
reference toxicant(s) are selected and used.
1257
1258
1259
4.17
1260
RECORD KEEPING
1261
1262
1263
1264
1265
4.17.1
1266
Proper record keeping is important. A complete file
1267
should be maintained for each individual toxicity test or group of
1268
tests on closely related samples. This file should contain a record
1269
of the sample chain-of-custody; a copy of the sample log sheet; the
1270
original bench sheets for the test organism responses during the
1271
toxicity test(s); chemical analysis data on the sample(s); detailed
1272
records of the test organisms used in the test(s), such as species,
1273
source, age, date of receipt, and other pertinent information
1274
relating to their history and health; information on the
1275
calibration of equipment and instruments; test conditions employed;
1276
and results of reference toxicant tests. Laboratory data should be
1277
recorded on a real-time basis to prevent the loss of information or
1278
inadvertent introduction of errors into the record. Original data
1279
sheets should be signed and dated by the laboratory personnel
1280
performing the tests.
1281
1282
1283
4.17.2
1284
The regulatory authority should retain records pertaining
1285
to discharge permits. Permittees are required to retain records
1286
pertaining to permit applications and compliance for a minimum of 3
1287
years [40 CFR 122.41(j)(2)].
1288
1289
1290
FACILITIES AND EQUIPMENT
1291
5.1 GENERAL REQUIREMENTS
1292
1293
1294
5.1.1
1295
Effluent toxicity tests may be performed in a fixed or
1296
mobile laboratory. Facilities should include equipment for rearing
1297
and/or holding organisms.
1298
1299
1300
5.1.2
1301
The facilities must be well ventilated and free of toxic
1302
fumes. Sample preparation, culturing, and toxicity testing areas
1303
should be separated to avoid cross contamination of cultures or
1304
toxicity test solutions with toxic fumes. Laboratory ventilation
1305
systems should be checked to ensure that return air from chemistry
1306
laboratories and/or sample handling areas is not circulated to test
1307
organism culture rooms or toxicity test rooms, or that air from
1308
toxicity test rooms does not contaminate culture areas. Air
1309
pressure differentials between such rooms should not result in a
1310
net flow of potentially contaminated air to sensitive areas through
1311
open or loosely-fitting doors.
1312
1313
1314
5.1.3
1315
Control of test solution temperature can best be achieved
1316
using circulating water baths, heat exchangers, or environmental
1317
chambers. Photoperiod can be controlled using automatic timers in
1318
the laboratory or environmental chambers.
1319
1320
1321
5.1.4
1322
Water used for rearing, holding, and testing organisms
1323
may be reconstituted synthetic water, ground water, surface water,
1324
or dechlorinated tap water. Dechlorination can be accomplished by
1325
carbon filtration, laboratory water conditioning units, or the use
1326
of sodium thiosulfate. After dechlorination, total residual
1327
chlorine should be non-detectable. Sodium thiosulfate may be toxic
1328
to the test organisms, and if used for dechlorination, paired
1329
controls with and without sodium thiosulfate should be incorporated
1330
in effluent toxicity tests. Use of 3.6 mg (anhydrous) sodium
1331
thiosulfate/L will reduce l.0 mg chlorine/L. After dechlorination,
1332
total residual chlorine should be non-detectable.
1333
1334
1335
1336
1337
5.1.4.1
1338
A good quality, laboratory grade deionized water,
1339
providing a resistance of 18 megaohm-cm, must be available in the
1340
laboratory and in sufficient quantity for laboratory needs.
1341
Deionized water may be obtained from MILLIPORE®, Milli-Q®,
1342
MILLIPORE QPAKTM2 or equivalent system. If large quantities of high
1343
quality deionized water are needed, it may be advisable to supply
1344
the laboratory grade water deionizer with preconditioned water from
1345
a CULLIGAN®, CONTINENTAL®, or equivalent, mixed-bed water treatment
1346
system.
1347
1348
1349
5.1.5
1350
Air used for aeration must be free of oil and fumes.
1351
Oil-free air pumps should be used where possible. Particulates can
1352
be removed from the air using BALSTON® Grade BX or equivalent
1353
filters (Balston, Inc., Lexington, MA), and oil and other organic
1354
vapors can be removed using activated carbon filters (BALSTON®, C-1
1355
filter, or equivalent).
1356
1357
1358
5.1.6
1359
During rearing, holding, and testing, test organisms
1360
should be shielded from external disturbances such as rapidly
1361
changing light conditions (especially salmonids) and pedestrian
1362
traffic.
1363
1364
1365
5.1.7
1366
Materials used for exposure chambers, tubing, etc., that
1367
come in contact with the effluent and dilution water should be
1368
carefully chosen. Tempered glass and perfluorocarbon plastics
1369
(TEFLON®) should be used whenever possible to minimize sorption and
1370
leaching of toxic substances, and may be reused after cleaning.
1371
Containers made of plastics, such as polyethylene, polypropylene,
1372
polyvinyl chloride, TYGON®, etc., may be used to ship, store, and
1373
transfer effluents and receiving waters, but they should not be
1374
reused unless absolutely necessary, because they could carry over
1375
adsorbed toxicants from one test to another. However, these
1376
containers may be repeatedly reused for storing uncontaminated
1377
waters such as deionized or laboratory-prepared dilution waters and
1378
receiving waters. Glass or disposable polystyrene containers can be
1379
used as test chambers. The use of large ($20 L) glass carboys is
1380
discouraged for safety reasons.
1381
1382
1383
5.1.8
1384
New plastic products should be tested for toxicity before
1385
general use by exposing organisms to them under ordinary test
1386
conditions.
1387
1388
1389
5.1.9
1390
Equipment which cannot be discarded after each use
1391
because of cost, must be decontaminated according to the cleaning
1392
procedures listed below. Fiberglass, in addition to the previously
1393
mentioned materials, can be used for holding and dilution water
1394
storage tanks, and in the water delivery system. All material
1395
should be flushed or rinsed thoroughly with dilution water before
1396
using in the test.
1397
1398
1399
1400
1401
5.1.10
1402
Copper, galvanized material, rubber, brass, and lead must
1403
not come in contact with holding or dilution water, or with
1404
effluent samples and test solutions. Some materials, such as
1405
neoprene rubber (commonly used for stoppers), may be toxic and
1406
should be tested before use.
1407
1408
1409
5.1.11
1410
Silicone adhesive used to construct glass test chambers
1411
absorbs some organochlorine and organophosphorus pesticides, which
1412
are difficult to remove. Therefore, as little of the adhesive as
1413
possible should be in contact with water. Extra beads of adhesive
1414
inside the containers should be removed.
1415
1416
1417
5.2
1418
CLEANING TEST CHAMBERS AND LABORATORY
1419
APPARATUS
1420
1421
1422
1423
1424
5.2.1
1425
New plasticware used for effluent or dilution water
1426
collection or organism test chambers does not require thorough
1427
cleaning before use. It is sufficient to rinse new sample
1428
containers once with sample dilution water before use. New
1429
glassware must be soaked overnight in 10% acid (see below) and
1430
rinsed well in deionized water and dilution water.
1431
1432
1433
1434
5.2.2
1435
All non-disposable sample containers, test vessels,
1436
tanks, and other equipment that has come in contact with effluent
1437
must be washed after use in the manner described below to remove
1438
surface contaminants as described below:
1439
1440
1441
1442
1.
1443
Soak 15 min in tap water, and scrub with detergent, or
1444
clean in an automatic dishwasher.
1445
1446
1447
2.
1448
Rinse twice with tap water.
1449
1450
1451
3.
1452
Carefully rinse once with fresh, dilute (10%, V:V)
1453
hydrochloric or nitric acid to remove scale, metals, and bases. To
1454
prepare a 10% solution of acid, add 10 mL of concentrated acid to
1455
90 mL of deionized water.
1456
1457
1458
4.
1459
Rinse twice with deionized water.
1460
1461
1462
5.
1463
Rinse once with full-strength, pesticide-grade acetone to
1464
remove organic compounds (use a fume hood or canopy).
1465
1466
1467
6.
1468
Rinse three times with deionized water.
1469
1470
1471
1472
1473
1474
1475
5.2.3
1476
All test chambers and equipment should be thoroughly
1477
rinsed with the dilution water immediately prior to use in each
1478
test.
1479
1480
1481
5.3
1482
APPARATUS AND EQUIPMENT FOR CULTURING AND TOXICITY
1483
TESTS
1484
1485
1486
1487
1488
5.3.1
1489
Culture units -- see Appendix. It is preferable to obtain
1490
test organisms from in-house culture units. If it is not feasible
1491
to maintain cultures in-house, test organisms can be obtained from
1492
commercial sources, and should be shipped to the laboratory in well
1493
oxygenated water in insulated containers to minimize excursions in
1494
water temperature during shipment. The temperature of the water in
1495
the shipping containers should be measured on arrival, to determine
1496
if the organisms were subjected to obvious undue thermal
1497
stress.
1498
1499
1500
5.3.2
1501
Samplers -- automatic samplers, preferably with sample
1502
cooling capability, that can collect a 24-h composite sample of 2 L
1503
or more.
1504
1505
1506
5.3.3
1507
Sample containers -- for sample shipment and storage (see
1508
Section 8, Effluent and Receiving Water Sampling and Sample
1509
Handling).
1510
1511
1512
5.3.4
1513
Environmental chamber or equivalent facility with
1514
temperature control (20EC or 25EC)
1515
1516
1517
5.3.5
1518
Water purification system -- MILLIPORE® MILLI-Q®,
1519
MILLIPORE® QPAK™2, or equivalent. Depending on the quantity of high
1520
grade water needed, a first-stage pre-conditioner deionizer, such
1521
as a Culligan® or Continental® System, or equivalent, may be needed
1522
to provide feed water to the high-purity system.
1523
1524
1525
5.3.6
1526
Balance -- analytical, capable of accurately weighing to
1527
0.0001 g.
1528
1529
1530
5.3.7
1531
Reference weights, Class S -- for documenting the
1532
performance of the analytical balance(s). The balance(s) should be
1533
checked with reference weights which are at the upper and lower
1534
ends of the range of the weighings made when the balance is used. A
1535
balance should be checked at the beginning of each series of
1536
weighings, periodically (such as every tenth weight) during a long
1537
series of weighings, and after the last weight of a series is
1538
taken.
1539
1540
1541
5.3.8
1542
Test chambers -- borosilicate glass or non-toxic
1543
disposable plastic test chambers are suitable. Test chamber volumes
1544
are indicated in the method summaries. To avoid potential
1545
contamination from the air and excessive evaporation of test
1546
solutions during the test, the chambers should be covered with
1547
safety glass plates or sheet plastic, 6 mm (¼ in) thick.
1548
1549
1550
5.3.9
1551
Volumetric flasks and graduated cylinders -- Class A,
1552
borosilicate glass or non-toxic plastic labware, 10-1000 mL for
1553
making test solutions.
1554
1555
1556
1557
1558
5.3.10
1559
Volumetric pipets -- Class A, 1-100 mL.
1560
1561
1562
5.3.11
1563
Serological pipets -- 1-10 mL, graduated.
1564
1565
1566
5.3.12
1567
Pipet bulbs and fillers -- PROPIPET®, or
1568
equivalent.
1569
1570
1571
5.3.13
1572
Droppers, and glass tubing with fire polished edges, 4 mm
1573
ID -- for transferring test organisms.
1574
1575
1576
5.3.14
1577
Wash bottles -- for rinsing small glassware and
1578
instrument electrodes and probes.
1579
1580
1581
5.3.15
1582
Glass or electronic thermometers -- for measuring water
1583
temperature.
1584
1585
1586
5.3.16
1587
Bulb-thermograph or electronic-chart type thermometers --
1588
for continuously recording temperature.
1589
1590
1591
5.3.17
1592
National Bureau of Standards Certified thermometer (see
1593
USEPA Method 170.1; USEPA 1979b).
1594
1595
1596
5.3.18
1597
pH, DO, and specific conductivity meters -- for routine
1598
physical and chemical measurements. Unless the test is being
1599
conducted to specifically measure the effect of one of the above
1600
parameters, a portable, field-grade instrument is
1601
acceptable.
1602
1603
1604
5.3.19
1605
Refractometer -- for measuring effluent, receiving, and
1606
test solution salinity.
1607
1608
1609
5.3.20
1610
Amperometric titrator -- for measuring total residual
1611
chlorine.
1612
1613
1614
5.4
1615
REAGENTS AND CONSUMABLE MATERIALS
1616
1617
1618
1619
1620
5.4.1
1621
Reagent water -- defined as MILLIPORE® MILLI-Q®,
1622
MILLIPORE® QPAK™2 or equivalent water (see Subsection 5.3.5
1623
above).
1624
1625
1626
5.4.2
1627
Effluent, dilution water, and receiving water -- see
1628
Section 7, Dilution Water, and Section 8, Effluent and Receiving
1629
Water Sampling and Sample Handling.
1630
1631
1632
5.4.3
1633
Reagents for hardness and alkalinity tests (see USEPA
1634
Methods 130.2 and 310.1; USEPA l979b).
1635
1636
1637
5.4.4
1638
Standard pH buffers 4, 7, and 10 (or as per instructions
1639
of instrument manufacturer) for instrument calibration (see USEPA
1640
Method 150.1; USEPA 1979b).
1641
1642
1643
5.4.5
1644
Specific conductivity and salinity standards (see USEPA
1645
Method 120.1; USEPA 1979b).
1646
1647
1648
5.4.6
1649
Laboratory quality control check samples and standards
1650
for the above chemistry methods.
1651
1652
1653
5.4.7
1654
Reference toxicant solutions (see Section 4, Quality
1655
Assurance).
1656
1657
1658
5.4.8
1659
Membranes and filling solutions for dissolved oxygen
1660
probe (see USEPA Method 360.1; USEPA 1979b), or reagents for
1661
modified Winkler analysis.
1662
1663
1664
5.4.9
1665
Sources of Food for Cultures and Toxicity
1666
Tests.
1667
1668
1669
1670
1671
5.4.9.1
1672
All food should be tested for nutritional suitability,
1673
and chemically analyzed for organic chlorine, PCBs, and toxic
1674
metals (see Section 4, Quality Assurance).
1675
1676
1677
1678
5.4.9.2
1679
Brine Shrimp (Artemia) -- see Appendix A.
1680
1681
1682
1683
1684
1.
1685
Brine Shrimp (Artemia) Cysts.
1686
1687
There are many commercial sources of brine shrimp cysts. The
1688
quality of the cysts may vary from one batch to another, and the
1689
cysts in each new batch (can or lot) should be evaluated for
1690
nutritional suitability and chemical contamination. The nutritional
1691
suitability (see Leger et al., 1985, 1986) of each new batch is
1692
checked against known suitable reference cysts by performing a
1693
side-by-side growth and/or reproduction tests using the "new" and
1694
"reference" cysts. If the results of tests for nutritional
1695
suitability or chemical contamination do not meet standards, the
1696
Artemia should not be used.
1697
1698
1699
2.
1700
Frozen Adult Brine Shrimp Frozen adult brine shrimp are
1701
available from pet stores and other commercial sources.
1702
1703
1704
1705
1706
5.4.9.3 Trout Chow
1707
Starter or No. 1 pellets, prepared according to current U.S.
1708
Fish and Wildlife Service specifications, are available from
1709
commercial sources. (The flake food, TETRAMIN® or BIORIL®, can be
1710
used regularly as a substitute for trout chow in preparing food for
1711
daphnids, and can be used as a short-term substitute for trout chow
1712
in feeding fathead minnows.)
1713
5.4.9.4 Dried, Powdered Leaves (CEROPHYLL®)
1714
Dried, powdered, cereal leaves (e.g., CEROPHYLL® or equivalent)
1715
are available from commercial suppliers. Dried, powdered, alfalfa
1716
leaves obtained from health food stores have been found to be a
1717
satisfactory substitute for cereal leaves.
1718
1719
1720
5.4.9.5
1721
Yeast Packaged dry yeast, such as Fleischmann's, or
1722
equivalent, can be purchased at the local grocery store.
1723
1724
1725
5.4.9.6
1726
Flake Fish Food The flake foods, TETRAMIN® and BIORIL®,
1727
are available at most pet supply shops.
1728
1729
1730
5.5 TEST ORGANISMS
1731
5.5.1 Test organisms are obtained from inhouse cultures or
1732
commercial suppliers (see Section 6, Test Organisms).
1733
TEST ORGANISMS
1734
6.1 TEST SPECIES
1735
1736
1737
6.1.1
1738
The species used in characterizing the acute toxicity of
1739
effluents and/or receiving waters will depend on the requirements
1740
of the regulatory authority and the objectives of the test. It is
1741
essential that good quality test organisms be readily available
1742
throughout the year from inhouse or commercial sources to meet
1743
NPDES monitoring requirements. The organisms used in toxicity tests
1744
must be identified to species. If there is any doubt as to the
1745
identity of the test organisms, representative specimens should be
1746
sent to a taxonomic expert to confirm the
1747
identification.
1748
1749
1750
6.1.2
1751
Toxicity test conditions and culture methods are provided
1752
in this manual for the following principal test
1753
organisms:
1754
1755
1756
Freshwater Organisms:
1757
1758
1759
1.
1760
Ceriodaphnia dubia (daphnid) (Table 12).
1761
1762
1763
2.
1764
Daphnia pulex and D. magna (daphnids) (Table
1765
13).
1766
1767
1768
3.
1769
Pimephales promelas (fathead minnow) (Table
1770
14).
1771
1772
1773
4.
1774
Oncorhynchus mykiss (rainbow trout) and Salvelinus
1775
fontinalis (brook trout) (Table 15).
1776
1777
1778
Estuarine and Marine Organisms:
1779
1780
1781
1.
1782
Mysidopsis bahia (mysid) (Table 16).1
1783
1784
1785
2.
1786
Cyprinodon variegatus (sheepshead minnow) (Table
1787
17).
1788
1789
1790
3.
1791
Menidia beryllina (inland silverside), M. menidia
1792
(Atlantic silverside), and M. peninsulae (tidewater silverside)
1793
(Table 18).
1794
1795
1796
1797
1798
6.1.3
1799
The test species (AFS, 1991) listed in Subsection 6.1.2
1800
are the recommended acute toxicity test organisms. They are easily
1801
cultured in the laboratory, are sensitive to a variety of
1802
pollutants, and are generally available throughout the year from
1803
commercial sources. Summaries of test conditions for these species
1804
are provided in Tables 12-18. Guidelines for culturing and/or
1805
holding the organisms are provided in Appendix A.
1806
1807
1808
6.1.4
1809
Additional species may be suitable for toxicity tests in
1810
the NPDES Program. A list of alternative acute toxicity test
1811
species and minimal testing requirements (i.e., temperature,
1812
salinity, and life stage) for these species are provided in
1813
Appendix B. Table 19 provides a summary of test conditions for
1814
Holmesimysis costata, which should also be considered an
1815
alternative acute toxicity test species. The Holmesimysis costata
1816
Acute Test (Table 19) is specific to Pacific Coast waters and is
1817
not listed at 40 CFR Part 136 for nationwide use. It is important
1818
to note that these species may not be as easily cultured or tested
1819
as the species on the list in 6.1.2, and may not be available from
1820
commercial sources.
1821
1822
1823
6.1.5
1824
Some states have developed culturing and testing methods
1825
for indigenous species that may be as sensitive or more sensitive
1826
than the species recommended in 6.1.2. However, EPA allows the use
1827
of indigenous species only where state regulations require their
1828
use or prohibit importation of the species in 6.1.2. Where state
1829
regulations prohibit importation or use of the recommended test
1830
species, permission must be requested from the appropriate state
1831
agency prior to their use.
1832
1833
1834
1 The genus name of this organism was formally changed to
1835
Americamysis (Price et al., 1994); however, the method manual will
1836
continue to refer to Mysidopsis bahia to maintain consistency with
1837
previous versions of the method.
1838
1839
1840
6.1.6
1841
Where states have developed culturing and testing methods
1842
for indigenous species other than those recommended in this manual,
1843
data comparing the sensitivity of the substitute species and one or
1844
more of the recommended species must be obtained in side-by-side
1845
toxicity tests with reference toxicants and/or effluents, to ensure
1846
that the species selected are at least as sensitive as the
1847
recommended species. These data must be submitted to the permitting
1848
authority (State or Region) if required. EPA acknowledges that
1849
reference toxicants prepared from pure chemicals may not always be
1850
representative of effluents. However, because of the observed
1851
and/or potential variability in the quality and toxicity of
1852
effluents, it is not possible to specify a representative
1853
effluent.
1854
1855
1856
1857
6.1.7
1858
Guidance for the selection of test organisms where the
1859
salinity of the effluent and/or receiving water requires special
1860
consideration is provided in the Technical Support Document for
1861
Water Quality-Based Toxics Control (USEPA, 1991c).
1862
1863
1864
1865
1.
1866
Where the salinity of the receiving water is <1‰,
1867
freshwater organisms are used regardless of the salinity of the
1868
effluent.
1869
1870
1871
2.
1872
Where the salinity of the receiving water is $1‰, the
1873
choice of organisms depends on state water quality standards and/or
1874
permit requirements.
1875
1876
1877
1878
1879
6.2
1880
SOURCES OF TEST ORGANISMS
1881
1882
1883
6.2.1 INHOUSE CULTURES
1884
1885
1886
6.2.1.1
1887
Inhouse cultures should be established wherever it is
1888
cost effective. If inhouse cultures cannot be maintained, test
1889
organisms should be purchased from experienced commercial suppliers
1890
(see Appendix for sources).
1891
1892
1893
6.2.2
1894
COMMERCIAL SUPPLIERS
1895
1896
1897
1898
1899
6.2.2.1
1900
All of the principal test organisms listed in Subsection
1901
6.1.2 are available from commercial suppliers.
1902
1903
1904
6.2.3
1905
FERAL (NATURAL OCCURRING, WILD CAUGHT)
1906
ORGANISMS
1907
1908
1909
6.2.3.1 The use of test organisms taken from the receiving water
1910
has strong appeal, and would seem to be the logical approach.
1911
However, it is impractical for the following reasons:
1912
1913
1914
1.
1915
Sensitive organisms may not be present in the receiving
1916
water because of previous exposure to the effluent or other
1917
pollutants.
1918
1919
1920
2.
1921
It is often difficult to collect organisms of the
1922
required age and quality from the receiving water;
1923
1924
1925
3.
1926
Most states require collection permits, which may be
1927
difficult to obtain. Therefore, it is usually more cost effective
1928
to culture the organisms in the laboratory or obtain them from
1929
private, state, or Federal sources. Fish such as fathead minnows,
1930
sheepshead minnows, and silversides, and invertebrates such as
1931
daphnids and mysids, are easily reared in the laboratory or
1932
purchased.
1933
1934
1935
4.
1936
The required QA/QC records, such as the single laboratory
1937
precision data, would not be available.
1938
1939
1940
5.
1941
Since it is mandatory that the identity of test organisms
1942
is known to the species level, it would necessary to examine each
1943
organism caught in the wild to confirm its identity, which would
1944
usually be impractical or, at the least, very stressful to the
1945
organisms.
1946
1947
1948
6.
1949
Test organisms obtained from the wild must be observed in
1950
the laboratory for a minimum of one week prior to use, to assure
1951
that they are free of signs of parasitic or bacterial infections
1952
and other adverse effects. Fish captured by electroshocking must
1953
not be used in toxicity testing.
1954
1955
1956
1957
1958
6.2.3.2
1959
Guidelines for collection of feral organisms are provided
1960
in USEPA, 1973; USEPA 1990a.
1961
1962
1963
6.2.4
1964
Regardless of their source, test organisms should be
1965
carefully observed to ensure that they are free of signs of stress
1966
and disease, and in good physical condition. Some species of test
1967
organisms, such as trout, can be obtained from stocks certified as
1968
"disease-free."
1969
1970
1971
6.3
1972
LIFE STAGE
1973
1974
1975
1976
1977
6.3.1
1978
Young organisms are often more sensitive to toxicants
1979
than are adults. For this reason, the use of early life stages,
1980
such as first instars of daphnids and juvenile mysids and fish, is
1981
required for all tests. In a given test, all organisms should be
1982
approximately the same age and should be taken from the same
1983
source. Since age may affect the results of the tests, it would
1984
enhance the value and comparability of the data if the same species
1985
in the same life stages were used throughout a monitoring program
1986
at a given facility.
1987
1988
1989
6.4
1990
LABORATORY CULTURING
1991
1992
1993
1994
1995
6.4.1
1996
Instructions for culturing and/or holding the recommended
1997
test organisms are included in Appendix A.
1998
1999
2000
6.5
2001
HOLDING AND HANDLING TEST ORGANISMS
2002
2003
2004
2005
2006
6.5.1
2007
Test organisms should not be subjected to changes of more
2008
than 3EC in water temperature or 3‰ in salinity in any 12 h
2009
period.
2010
2011
2012
6.5.2
2013
Organisms should be handled as little as possible. When
2014
handling is necessary, it should be done as gently, carefully, and
2015
quickly as possible to minimize stress. Organisms that are dropped
2016
or touch dry surfaces or are injured during handling must be
2017
discarded. Dipnets are best for handling larger organisms. These
2018
nets are commercially available or can be made from small-mesh
2019
nylon netting, silk bolting cloth, plankton netting, or similar
2020
material. Wide-bore, smooth glass tubes (4 to 8 mm inside diameter)
2021
with rubber bulbs or pipettors (such as a PROPIPETTE® or other
2022
pipettor) should be used for transferring smaller organisms such as
2023
daphnids, mysids, and larval fish.
2024
2025
2026
6.5.3
2027
Holding tanks for fish are supplied with a good quality
2028
water (see Section 5, Facilities and Equipment) with a flow-through
2029
rate of at least two tank-volumes per day. Otherwise, use a
2030
recirculation system where the water flows through an activated
2031
carbon or undergravel filter to remove dissolved metabolites.
2032
Culture water can also be piped through high intensity ultraviolet
2033
light sources for disinfection, and to photodegrade dissolved
2034
organics.
2035
2036
2037
6.5.4
2038
Crowding should be avoided. The DO must be maintained at
2039
a minimum of 4.0 mg/L for marine and warm water, freshwater
2040
species, and 6.0 mg/L for cold-water, freshwater species. The
2041
solubility of oxygen depends on temperature, salinity, and
2042
altitude. Aerate if necessary.
2043
2044
2045
6.5.5
2046
Fish should be fed as much as they will eat at least once
2047
a day with live or frozen brine shrimp or dry food (frozen food
2048
should be completely thawed before use). Brine shrimp can be
2049
supplemented with commercially prepared food such as Tetramin® or
2050
BioRil® flake food, or equivalent. Excess food and fecal material
2051
should be removed from the bottom of the tanks at least twice a
2052
week by siphoning.
2053
2054
2055
6.5.6
2056
Fish should be observed carefully each day for signs of
2057
disease, stress, physical damage, and mortality. Dead and abnormal
2058
specimens should be removed as soon as observed. It is not uncommon
2059
to have some fish (5-10%) mortality during the first 48 h in a
2060
holding tank because of individuals that refuse to feed on
2061
artificial food and die of starvation.
2062
2063
2064
6.5.7
2065
A daily record of feeding, behavioral observations, and
2066
mortality should be maintained.
2067
2068
2069
6.6
2070
TRANSPORTATION TO THE TEST SITE
2071
2072
2073
2074
2075
6.6.1
2076
Organisms are transported from the base or supply
2077
laboratory to a remote test site in culture water or standard
2078
dilution water in plastic bags or large-mouth screw-cap (500 mL)
2079
plastic bottles in styrofoam coolers. Adequate DO is maintained by
2080
replacing the air above the water in the bags with oxygen from a
2081
compressed gas cylinder, and sealing the bags. Another method
2082
commonly used to maintain sufficient DO during shipment is to
2083
aerate with an airstone which is supplied from a portable pump. The
2084
DO concentration must not fall below 4.0 mg/L for marine and
2085
warm-water, freshwater species, and 6.0 mg/L for cold-water,
2086
freshwater species.
2087
2088
2089
6.6.2
2090
Upon arrival at the test site, organisms are transferred
2091
to receiving water if receiving water is to be used as the test
2092
dilution water. All but a small volume of the holding water
2093
(approximately 5%) is removed by siphoning, and replaced slowly
2094
over a 10 to 15 min period with dilution water. If receiving water
2095
is used as dilution water, caution must be exercised in exposing
2096
the test organisms to it, because of the possibility that it might
2097
be toxic. For this reason, it is recommended that only
2098
approximately 10% of the test organisms be exposed initially to the
2099
dilution water. If this group does not show excessive mortality or
2100
obvious signs of stress in a few hours, the remainder of the test
2101
organisms are transferred to the dilution water.
2102
2103
2104
6.6.3
2105
A group of organisms must not be used for a test if they
2106
appear to be unhealthy, discolored, or otherwise stressed, or if
2107
mortality appears to exceed 10% preceding the test. If the
2108
organisms fail to meet these criteria, the entire group must be
2109
discarded and a new group obtained. The mortality may be due to the
2110
presence of toxicity, if receiving water is used as dilution water,
2111
rather than a diseased condition of the test organisms. If the
2112
acclimation process is repeated with a new group of test organisms
2113
and excessive mortality occurs, it is recommended that an
2114
alternative source of dilution water be used.
2115
2116
2117
6.6.4
2118
In static tests, marine organisms can be used at all
2119
concentrations of effluent by adjusting the salinity of the
2120
effluent to a standard salinity (such as 25‰) or to the salinity
2121
approximating that of the receiving water, by adding sufficient dry
2122
ocean salts, such as Forty Fathoms®, or equivalent, GP2 or
2123
hypersaline brine.
2124
2125
2126
6.6.5
2127
Saline dilution water can be prepared with deionized
2128
water or a freshwater such as well water or a suitable surface
2129
water. If dry ocean salts are used, care must be taken to ensure
2130
that the added salts are completely dissolved and the solution is
2131
aerated 24 h before the test organisms are placed in the solutions.
2132
The test organisms should be acclimated in synthetic saline water
2133
prepared with the dry salts. Caution: addition of dry ocean salts
2134
to dilution water may result in an increase in pH. (The pH of
2135
estuarine and coastal saline waters is normally
2136
7.5-8.3.)
2137
2138
2139
6.6.6
2140
All effluent concentrations and the control(s) used in a
2141
test should have the same salinity. However, if this is impractical
2142
because of the large volumes of water required, such as in
2143
flow-through tests, the highest effluent concentration (lowest
2144
salinity) that could be tested would depend upon the salinity of
2145
the receiving water and the tolerance of the test organisms. The
2146
required salinities for toxicity tests with estuarine and marine
2147
species are listed in Tables 16-19. However, the tolerances of
2148
other candidate test species would have to be determined by the
2149
investigator in advance of the test.
2150
2151
2152
6.6.7
2153
Because of the circumstances described above, when
2154
performing flow-through tests of effluents discharged to saline
2155
waters, it is advisable to acclimate groups of test organisms to
2156
each of three different salinities, such as 10, 20, and 30‰, prior
2157
to transporting them to the test site. It may also be advisable to
2158
maintain cultures of these test organisms at a series of salinity
2159
levels, including at least 10, 20, and 30‰, so that the change in
2160
salinity upon acclimation at the desired test dilutions does not
2161
exceed 6‰.
2162
2163
2164
6.7
2165
TEST ORGANISM DISPOSAL
2166
2167
2168
6.7.1 When the toxicity test is concluded, all test organisms
2169
(including controls) should be humanely destroyed and disposed of
2170
in an appropriate manner.
2171
2172
2173
2174
2175